작동 원리 및 구조
Run the pump into a measuring cylinder for a known time and enter the volume. The tool gives the real flow in mL/min, L/h and L/day and the correction against the label. Switch to dose mode to calculate how many millilitres of product reach a target ppm in a given volume and how long the pump must run.
단계별 사용 방법
Collect the pump output for a timed run at the working back-pressure.
Enter the volume and the time.
Compare the flow with the rated value.
Use dose mode to plan product volume and run time.
주요 기능 및 장점
- Real flow in several units
- Deviation from rated flow with warning
- ppm dose to product volume
- Pump run time
계산 방식, 공식 및 기술 표준
Flow = volume ÷ time × 60. Product (mL) = system litres × ppm ÷ strength (g/L). Run time = product ÷ flow.
입출력 예시 및 엣지 케이스
48 mL in 60 s
Calibration
48 mL/min, 2.88 L/h
제한 사항 및 정확도 안내
- • Flow changes with pressure, viscosity and tubing wear.
- • Check chemical compatibility and safety data before dosing.
브라우저 로컬 처리 및 개인정보 보호
일반 웹 변환기는 문서나 미디어 파일을 원격 서버에 업로드합니다. 정량 펌프 보정 및 유량 계산기 도구는 현재 브라우저 탭 내부에서만 실행됩니다:
입력 텍스트와 파일은 외부 서버로 전송되지 않고 기기 내에서 로컬로 계산됩니다.
도구에 붙여넣은 파일이나 데이터를 서버에 기록하거나 저장하지 않습니다.
작업 결과는 브라우저 탭 메모리에만 유지되며 탭을 닫거나 초기화하면 즉시 해제됩니다.
자주 묻는 질문 (FAQ)
How often should I recalibrate?
Whenever you change tubing, product or pressure, and regularly for critical dosing.
What is 1 ppm?
One milligram per litre in water.
Is my data sent anywhere?
No.